- Pneumatic, measures pressure and vacuum
- Generates Pressure/Vacuum -600 mbar to +2000 mbar (built-in electrical pump)
- Pressure ranges to 1 mbar, 10 mbar, 100 mbar, 1000 mbar and 2000 mbar
- Accuracy up to ±0.1% FS (+1 digit)
- Great adjustment accuracy 0.01% FS
- Chargeable Li-Ion battery
- Generates pressure/vacuum by pressing a key
Type | LCC 100-1 | LCC 100-10 | LCC 100-100 | LCC 100-1000 | LCC 100-2000 |
Order-Code | LCC-100-1 | LCC-100-10 | LCC-100-100 | LCC-100-1000 | LCC-100-2000 |
Pressure range | -1...+1 mbar | -10...+10 mbar | -100...+100 mbar | -1000...+1000 mbar | -1000...+2000 mbar |
Overpressure | 5-fold | 5-fold | 5-fold | 2-fold | 2-fold |
Measurement Uncertainty (+1 digit) | ±0.3% FS | ±0.1% FS | ±0.1% FS | ±0.1% FS | ±0.1% FS |
Linearity ±1digit | ±0.2% FS | ±0.1% FS | ±0.1% FS | ±0.1% FS | ±0.1% FS |
Hysteresis | max. 0.1% FS | ||||
Measurement principle | inductive | ||||
Temperature drift of the internal reference sensor | Zero: 0.003% FS / K (0% via zero balance) Span: 0.03% FS / K | ||||
Zero balance | • automatic (in settable time spans, possible to swith of), or • manuel (by pressing the ZERO button) | ||||
Long-term stability of the internal reference sensor | 0.1% FS per year (typical) | ||||
Temperature ranges | Working range +10...+40°C; Storage -10...+70°C | ||||
Usable pressure measuring range | -10...+110% | ||||
Adjustment accuracy | 0.01% FS | ||||
Adjustment time | depending on volume, <5 sec | ||||
Measurement media | Air, non-aggressive and corrosion-free dry gases | ||||
Operation modes | • CTRL (controlling pressure) • MESS (measuring pressure) • AUTO (individually definable pressure profile) • Remote-controlled operation (via USB or RS232 interface) | ||||
Display | Graphic display (white on dark blackground) | ||||
Interfaces | USB and RS232 | ||||
Supply | 24 VDC / 1 A, build-in Li-Ion battery (8h typical) | ||||
Pressure ports | 6.6 x 11 mm (for flexible hose with 6 mm inner diameter) | ||||
Dimension | H 102,6 x W 257 x D 271 mm (without handel) |