Basic Specifications: | Withstanding Test Section |
Testing Voltage: | 0.2 kV to 5.00 kV AC, 500VA (maximum 30 minutes) |
Voltage Setting Method: | Digital Setting |
Waveform: | Same as the power supply waveform |
Frequency: | Same as the power supply frequency |
Measurement Range: | Current: 0.01 to 100.0 mA, ±(2 % f.s. +5 dgt.) 10 mA/100 mA AC (Average value rectified, effective value display) |
Basic Specifications: | Insulation Test Section |
Testing Voltage: | 50 V to 1000 V DC |
Measurement Range: | 0.1 M-Ohm to 9999 M-Ohm, 4 ranges |
Basic Specifications: | General |
Judgement Function: | Contents: UPPER-FAIL / PASS / LOWER-FAIL (Digital setting window comparator method) |
Timer Section: | Setting range: 0.3 to 999 seconds |
Interfaces: | EXT I/O, EXT SW, RS-232C, GP-IB |
Display: | Fluorescent tube display (digital), Analog meter |
Monitor Function: | Output voltage, detection current |
Power Supply: | 100 V to 120 V AC / 200 to240 V AC, (50/60 Hz), 1000 VA max. |
Dimensions: | 320 mm(12.60 in)W × 155 mm(6.10 in)H × 480 mm(18.90 in)D |
Weight: | 18kg (634.9 oz) |
Accessories: | 9615 H.V. TEST LEAD (high voltage side and return, 1 each), Power cord (1), spare fuse (1) |
Chi tiết
Programable testing, full remote control, Automatic Insulation Withstanding Tester |
This electrical safety test instrument series is designed for insulation resistance and voltage withstand testing of electrical devices and components according to various safety standards. A multitude of automation and labor-saving features are provided to ensure effective testing for a wide variety of requirements and test conditions. Select the most appropriate model for your applications.
|
Basic specifications : Withstanding test section |
| ||||||||||
Basic specifications : Insulation test section |
| ||||
Basic specifications : general |
Judgment function | Contents: UPPER-FAIL / PASS / LOWER-FAIL (Digital setting window comparator method) |
Timer section | Setting range: 0.3 to 999 seconds |
Interfaces | EXT I/O, EXT SW, RS-232C, GP-IB |
Display | Fluorescent tube display (digital), Analog meter |
Monitor function | Output voltage, detection current |
Power supply | 100 V to 120 V AC / 200 to240 V AC, (50/60 Hz), 1000 VA max. |
Dimensions, mass | 320 mm(12.60 in)W × 155 mm(6.10 in)H × 480 mm(18.90 in)D, 18 kg (634.9 oz) |
Accessories | H.V. TEST LEAD 9615 (high voltage side and return, 1 each), Power cord (1), spare fuse (1) |